GD-3536A Automatic Cleveland Open Cup Flash Point Analyzer

The Automatic Cleveland Open Cup Flash Point Tester adopts LCD screen to display and full English man-machine interface. It can preset expected flash point temperature, sample mark number, barometric pressure and other parameters. Automatically open the lid, detect the flash point and print test data.
Availability:
Quantity:
  • GD-3536A

  • GOLD

  • GD-3536A

GD-3536A Automatic Cleveland Open Cup Flash Point Tester



Introduction:
The Automatic Cleveland Open Cup Flash Point Tester is designed and made as per GB/T3536-2008 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. 

It is applicable to all petroleum products with flash points above 79℃ and below 400℃ except fuel oils.
 

Features:
1.  The Automatic Cleveland Open Cup Flash Point Tester adopts LCD screen to display and full English man-machine interface. It can preset expected flash point temperature, sample mark number, barometric pressure and other parameters.It has menu prompt and input guide。

2.  It adopts simulation tracking to display the function curve of temperature rising and test time. It has the functions of English mis-operation prompt,test date,test time and other parameters.

3.  Automatically calibrate the effect of atmospheric pressure to test and calculate the corrected value.

4.  Differential coefficient detection.Automatically correct the system deviation.

5.  Automatically open the lid,detect the flash point and print test data.The test arm automatically rise up and lower down.
6.  Reasonable design, safe and convenient to operate.
 

Specifications:

Power supplyAC220(-10%~+5%)V, 50Hz
Flash point determinationRange: Ambient to 400℃
Repeatability: ≤8℃
Reproducibility: ≤17℃
Accuracy: 0.1℃
Heating rateCorrespond with GB/T3536 and ASTM D92
Ignition modeElectric ignition. 
Gas flame diameter is 3.2mm~4.8mm
Ambient temperature(10~40)℃
Relative humidity≤80%
Maximum power consumption400W
Dimension410*360*310mm


Previous: 
Next: 
Product Inquiry

Gold Mechanical & Electrical Equipment

We're here to offer you professional, fast, comprehensive and high-quality services.
Our marketing network covers all countries worldwide and thus can provide high-quality pre-sale, in-sale and after-sale services for customers as soon as possible. 

Products

Quick Links

​Copyright © Chongqing Gold Mechanical&Electrical Equipment Co.,Ltd.  All Rights Reserved. Site Map  Supported by Leadong.com