GD-21789 ISO 13736 / IP 170 Automatic Abel Flash Point Apparatus
Reference Standards:
BS EN ISO 13736 Determination of flash point by Abel closed-cup method. It is for determination of the closed-cup flash point of combustible liquids having flash point between -30℃ and 70℃ -(22 and 158°F).
IP 170 Determination of flash point by Abel closed-cup method. It specifies a method for the determination of the closed-cup flash point of combustible liquids having flash points between -30.0℃ and 70.0℃. However, the precision given for this method is only valid for flash points in the range -5.0℃ to 66.5℃.
Introduction:
The Automated Abel Closed Cup Flash Point Apparatus is applied to the flammable and combustible materials for shipping and safety regulations. It consists of test flame applicator, brass test cup, thermometer support, heating plate and electric heater.
The Automatic Abel Closed Cup Flash Point Appratus is designed and manufactured in accordance with ISO 13736 / IP 170. It is suitable for the determination of automated closed-cup flash point of combustible liquids having flash points between -30.0℃ to 75.0℃ (Expected Flash Point -30.0℃ to 70.0℃). However, the precision given for this instrument is only valid for flash points in the range between -8.5℃ to 75.0℃.
The Abel Apparatus is suitable for flash point applications of jet fuels, solvents, chemicals, etc.
Key Features:
7 inches high resolution LCD color display, touch screen buttons.
English HMI, real-time display of input and output states of the instrument, intuitive and user-friendly.
Electronic ignition, electronic flame, no gas source required to ensure environment friendly and laboratory safety.
Automatic mechanism movements during and after tests.
High accuracy temperature sensor for test sample temperature measurements.
Automatic control throughout test procedures.
Flash point detector with short response time and high sensitivity.
Automatic atmospheric pressure measurement and flash point automatically corrected to standard atmospheric pressure.
Large memory capacity for 256 test results storage.
It features time setting and data correction functions.
Arising temperature limiting function, once temperature reaches threshold, instrument terminates test and raise warning.
Microprocessor program control, test results can be automatically identified, stored and printed.
Main Parameters:
1 | Applicable standard | ISO 13736, IP 170, GB/T 21789 |
2 | Operating voltage and power | AC220V, 50Hz, 1200W |
3 | Operating interface | 7 inches high resolution LCD color display, 800*480 pixels |
4 | Flash point test range | -30.0℃~75.0℃, precision measurement range: -8.5℃~75.0℃ |
5 | Sample heating rate | 0.75~1.25℃/min |
6 | Ignition duration | 0.5℃ |
7 | Temperature measurement resolution | 0.1℃ |
8 | Atmospheric pressure detection | digital atmospheric pressure sensor chip, accuracy: ±0.3kPa @25℃ |
9 | Refrigeration system | Danfoss compressor |
10 | Heating system | heating vessel |
11 | Printer | thermal printer, 36 characters |
12 | Ambient temperature | 5~30℃; Humidity: ≤75% |
Automatic Abel Flash Point Apparatus: