Installation method for Falling Correction device
1. Test holder and falling correction device:
(1) Test holder: The instrument can put the holder of sample ring,sample and ect.
(2) Falling correction device: it is the special device for falling correction correction.
2.Installation Method:
(1) Please put the falling correction device below the upper board, and the screw holes of the two columns are aligned with the two circular holes in the upper board.
(3) The falling correction device is fixed on the upper board with stainless steel screw. The assembly after installation.
If you want to unload, just screw down two screws.
Attention: When the test holder leaves the factory, the falling correction device has been installed. The user can decide whether to unload or mount the test holder according to the actual situation of the sample falling.
Softening point test bath liquid medium selection
Range | Liquid medium | Memo |
<35℃ (higher than 0℃) | Distilled water and glycol 1+1 mixture | Softening point range 0℃~35℃,select 1:1 distilled water and glycol mixture as the medium. |
0℃~35℃ | The mixture of Water and Glycerin 1+1(volume ratio) | |
35℃~80℃ | Distilled water or deionized water | The boiled water cooled to 5℃~27℃. |
80℃~150℃ | Glycerin | The water content of reused glycerin will be increased and influence the test results.If the any appearance change of glycerin, pls change new one. |
>150℃ | Silicone oil(50cSt viscosity) | If the any appearance change of silicone oil,pls change new one.Don’t use any silicone oil with gelatin,which means degradation. |