1.
When drainage operation is required, the specific method:
First cut off the power of the instrument, and empty the air in the pressure chamber, then remove the small cover with the observation window on the rear panel with a Phillips screwdriver, put the sewage outlet pipeline into the water receiving container, and finally turn the gray knob at the lower end of the air filter assembly counterclockwise to perform the drainage operation.
2.
Specific method:
first remove the upper panel of the instrument, then remove the fixing ring fixing the drying cylinder with a screwdriver, unplug the air pressure tubes at both ends of the drying cylinder, unscrew the sealing cover of the drying cylinder, and pour out the desiccant inside for replacement. Put the instrument in reverse order.