Working principle:
This instrument adopts industrial control computer system combining a colored touched type LCD to form an advanced full automatic micro carbon residue detection system. It can command every parts to run concertedly according to preset procedure. It can receive balance data; can automatically control the testing process according to standard requirements. It can make the temperature in coke chamber constant at 500℃±2℃ at a heating rate of 10℃~15℃. It can change the flow rate of nitrogen according to standard automatically. It can do each step according to timing sequence automatically and having alarms to prompt. Also the temperature rising curve will be shown on the screen timely. The system can collect the test data, calculate the test result automatically. It can save and print the data at operator’s will. This instrument is designed all English interface. All parameters and working conditions can be shown on the screen, which is convenient to use and easy to operate. The operator can operate the instrument normally with a few simple touch actions until the test report is displayed.
Main technical specifications and parameters:
1. Power supply: AC(220±10%)V, 50Hz
2. Power consumption: ≤1400W
3. Temperature of coke chamber: 500℃
4. Temperature control accuracy: ±2℃
5. Ambient temperature: 15℃~35℃
6. Relative humidity: ≤85%
7. Dimension: 500mm×280mm×550mm
8. NW: 18.5kg
Common faults and solutions:
Common Faults | Causes | Solutions |
There is no response after the instrument is powered on | 1. Power supply is not connected. 2. Overload protector works. | 1. Check and connect the working power supply. 2. Reset overload protector. |
Full set:
No. | Name | Unit | Qty | Remarks |
1 | Main unit | Set | 1 | |
2 | Sample tube holder | Piece | 1 | |
3 | Big sample tube(12×70mm) | Piece | 24 | |
4 | Small sample tube(12×35mm) | Piece | 24 | |
5 | Sample tube hook | Piece | 1 | |
6 | Air inlet pipe(ф6 PU pipe, 2m) | Set | 1 | Used to connect the steel cylinder with main machine |
7 | Exhaust pipe(PA12 nylon pipe, 2m) | Piece | 1 | Used to remove exhaust gas |
8 | Cover of the coke chamber | Piece | 1 | |
9 | Condensation collection bottle | Piece | 1 | |
10 | Keyboard and mouse | Set | 1 |
Touch Screen Carbon Residue Tester (Micromethod) for Petroleum Products
Working principle:
This instrument adopts industrial control computer system combining a colored touched type LCD to form an advanced full automatic micro carbon residue detection system. It can command every parts to run concertedly according to preset procedure. It can receive balance data; can automatically control the testing process according to standard requirements. It can make the temperature in coke chamber constant at 500℃±2℃ at a heating rate of 10℃~15℃. It can change the flow rate of nitrogen according to standard automatically. It can do each step according to timing sequence automatically and having alarms to prompt. Also the temperature rising curve will be shown on the screen timely. The system can collect the test data, calculate the test result automatically. It can save and print the data at operator’s will. This instrument is designed all English interface. All parameters and working conditions can be shown on the screen, which is convenient to use and easy to operate. The operator can operate the instrument normally with a few simple touch actions until the test report is displayed.
Main technical specifications and parameters:
1. Power supply: AC(220±10%)V, 50Hz
2. Power consumption: ≤1400W
3. Temperature of coke chamber: 500℃
4. Temperature control accuracy: ±2℃
5. Ambient temperature: 15℃~35℃
6. Relative humidity: ≤85%
7. Dimension: 500mm×280mm×550mm
8. NW: 18.5kg
Common faults and solutions:
Common Faults | Causes | Solutions |
There is no response after the instrument is powered on | 1. Power supply is not connected. 2. Overload protector works. | 1. Check and connect the working power supply. 2. Reset overload protector. |
Full set:
No. | Name | Unit | Qty | Remarks |
1 | Main unit | Set | 1 | |
2 | Sample tube holder | Piece | 1 | |
3 | Big sample tube(12×70mm) | Piece | 24 | |
4 | Small sample tube(12×35mm) | Piece | 24 | |
5 | Sample tube hook | Piece | 1 | |
6 | Air inlet pipe(ф6 PU pipe, 2m) | Set | 1 | Used to connect the steel cylinder with main machine |
7 | Exhaust pipe(PA12 nylon pipe, 2m) | Piece | 1 | Used to remove exhaust gas |
8 | Cover of the coke chamber | Piece | 1 | |
9 | Condensation collection bottle | Piece | 1 | |
10 | Keyboard and mouse | Set | 1 |